Method for calculation the absolute dose in the Monte Carlo simulation

Oanh Thi Luong 1
Luong Thanh Dang 1
Tai Thanh Duong 2, 3, *
  1. Nguyen Tat Thanh University
  2. Đồng Nai Hospital
  3. University of Science, VNU-HCM
Correspondence to: Tai Thanh Duong, Đồng Nai Hospital; University of Science, VNU-HCM. Email: pvphuc@vnuhcm.edu.vn.
Volume & Issue: Vol. 2 No. 5 (2018) | Page No.: 90-96 | DOI: 10.32508/stdjns.v2i5.783
Published: 2019-07-02

Online metrics


Statistics from the website

  • Abstract Views: 0
  • Galley Views: 0

Statistics from Dimensions

This article is published with open access by Viet Nam National University Ho Chi Minh City, Viet Nam. This article is distributed under the terms of the Creative Commons Attribution License (CC-BY 4.0) which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Abstract

In this study, we presented the method for calculation the absolute dose in the Monte Carlo simulation following the prescription of Popescu et al for the 6 MV photon energy. The BEAMnrc was used to simulate 6 MV photon beams from a Siemens Primus M5497 linear accelerator at DongNai general hospital. The DOSXYZnrc was then used to calculate the dose distribution in a homogeneous phantom (in form of CT images). The absolute dose obtained from the MC and TPS were compared with measured ones using an ion chamber (Farmer Type Chamber FC65-P, IBA). The average doses discrepancy between the simulated and measured dose was 0.53±0.37% and between the simulated and TPS was 1.00±0.51%. Results showed good agreement between simulated, measured and calculated dosed on a homogeneous phantom.

Comments